Installing the NoVent Microfluidics

Connecting the NoVent Microfludics Module to the TRACE 1600/1610
Thermo Scientific
f.
Push the capillary column through the detector base according to the detector in use.
g. Use the 1/4-in. wrench to tighten the retaining nut.
h. Use enough pressure necessary to obtain a good seal without overtighten.
This section provides instructions for installing the NoVent™ Microfluidics on your TRACE
1600 or TRACE 1610 GC, and your ISQ or TSQ 8000 Series mass spectrometer.
Contents
"Connecting the NoVent Microfludics Module to the TRACE 1600/1610"
page 399
"Installing the Mounting Bracket"
"Preparing the NoVent Microfluidics Restrictor Tubing"
"Attaching the Ferrule and Nut to the GC Column"
"Attaching the New Tubing to the Transfer Line"
"Connecting the Capillaries to the Microfluidics Splitter"
"Configuring the Post-Column"
"Using the Module"
According to the configuration of your TRACE 1600 or TRACE 1610 GC, a dummy
module is present in the free site where the detector module is not installed.
If no dummy module is present, and no vacant module position is present on either the
TRACE 1600 or TRACE 1610 GC or on the optional gas valve oven (if you have one), you
must take one of the existing and installed modules off in order to install the NoVent
Microfluidics.
on
page 402
on
page 416
on
page 418
8
Installing Optional Kits
Installing the NoVent Microfluidics
on
page 405
on
page 408
on
page 410
on
page 414
TRACE 1600/1610 Hardware Manual
on
399