Figure 502. Pulse Generator Safety Covers
1. Connect the high voltage discharge cable.
a. On the high voltage side, loosen the right nut that fix the pulse generator to the
bracket. Do not loosen the left nut, keep it tightened. Place the safety cover until the
nuts couple the slots on the bottom of the safety cover. See
Figure 503. High Voltage Discharge Cable Connection (1)
b. Tighten the right nut using a T30 torxhead screwdriver. See
Pulse Generator
Safety Cover
TRACE 1600/1610 Hardware Manual
9
Adding Modules
Adding a PDD Module
Safety Cover
Figure
503.
Figure
504.