Replacing the Encapsulated Flow Restrictors

Thermo Scientific
This section provides instruction for replacing the 1/4-in. OD x 1/4-in. length encapsulated
flow restrictors. The flow restrictors are located inside the manifold marked Gas Outputs on
the GDI electrical interface. See
Figure 515. Gas Outputs Manifold
To replace an encapsulated flow restrictor
1. Remove the Gas Outputs manifold from the GDI electrical interface by unscrewing its
fixing screws using a T20 Torxhead screwdriver. See
Figure 516. Gas Outputs Manifold
2. Remove the encapsulated flow restrictor of interest from its seat using tweezers.
See
Figure
517.
Figure 517. Flow Restrictor Replacement
Figure
515.
Figure
Encapsulated
Flow
Restrictors
9
Adding Modules
Adding a Generic Detector Interface
Gas Outputs
Manifold
516.
Pay attention to not lose
the O-ring placed into the
manifold
Manifold O-Rings
Pay attention to not
lose the internal
O-ring
Internal O-Rings
TRACE 1600/1610 Hardware Manual
511