Thermo Scientific
Figure 341. Adapter Installation on SSL Injector Bottom
Adapter for 1/8-in. OD Packed Column
a. Finger-tighten the retaining nut of the adapter until it start to grip the bottom of the
injector.
b. Use the 1/2 wrench to firmly tighten the retaining nut of the adapter with the base
seal from the bottom of the injector.
7. Remove the bottom parts of the FID, NPD, ECD, TCD, or FPD detector.
See
Figure
342.
Figure 342. Detector Bottom Parts Removal
Note Save the bottom parts of the detector in a safe place because will be reused
when you restore the original configuration for capillary columns.
a. Using the 1/2 wrench provided, unscrew the retaining nut from the bottom of the
detector.
Installing the Packed Column Adapters
Retaining Nut
Silver Seal
Adapter for 1/16-in. OD Packed Column
Detector Bottom
Retaining Nut
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits
361