Thermo Scientific TRACE 1600 Hardware Manual page 371

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Note During the extraction of the photomultiplier tube from its holder, the PMT
telescopic centering sleeve assemble could remain inside the holder or get out of it.
In this case, before reinserting the photomultiplier tube into its holder, the PMT
telescopic centering sleeve assemble must be correctly repositioned as described at the
step
9.
8. Replace the defective photomultiplier tube.
Note Wear clean, lint- and powder-free gloves when you handle the photomultiplier
tube and void exposing it to light.
a. Carefully extract the photomultiplier tube from its connector. See
Figure 319. Photomultiplier Tube Replacement (1)
b. Replace the photomultiplier tube with a new one. Make sure to correctly insert the
pins of the photomultiplier tube into the connector referring to the polarized
position. See
Figure
Figure 320. Connector Polarized Position
320.
Polarized Position
7
Detectors Advanced Maintenance
Replacing the FPD Photomultiplier Tube
Figure
319.
Connector
Photomultiplier Tube
TRACE 1600/1610 Hardware Manual
339