Cleaning Or Replacing The Fid Collecting Electrode - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

Cleaning or Replacing the FID Collecting Electrode

Thermo Scientific
Note For maintaining or replacing any other component not listed in this section, see
Chapter 7, "Detectors Advanced Maintenance."
"Removing/Replacing a Detector Module"
"Cleaning or Replacing the FID Jet"
Note Some of the cleaning procedures in this section require the use of methanol.
If methanol is unavailable or prohibited, substitute LCMS-grade or GC-grade ethanol
or isopropyl alcohol. Do not use denatured ethanol as it may contain impurities that
contaminate the GC-MS system.
To clean or replace the collecting electrode
1. Put the GC in standby condition.
2. Turn the flame off. The fuel gases, hydrogen and air, are automatically closed.
3. Cool the detector to room temperature.
4. Turn the make-up gas off.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
5. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
6. If external modules are present, unplug the power cable from the AC Input connector of
each module, and from the wall outlet.
7. Open the module flap cover.
8. Disconnect the signal, glow-plug, and polarizing cables from their contacts on the cell.
4
Performing Detectors Routine Maintenance
Maintaining a Flame Ionization Detector (FID)
on
page 277
on
page 287
TRACE 1600/1610 Hardware Manual
153

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents