Figure 246. FID Top Parts Removal
a. Unscrew and remove the detector cap, paying attention to not rotate the cell top
cover.
b. Remove the cell top cover and put it in a safe place. Do not damage the FID
collecting electrode pin.
10. Remove the electrode. See
a. Using forceps or tweezers, carefully extract the collecting electrode, the insulator ring
and the collector insulator through the top of the detector cell. Place all the parts
removed on a clean surface.
Figure
246.
7
Detectors Advanced Maintenance
Measuring the FID Gas Flows
Detector Cap
Cell Top Cover
Insulator Ring
Collecting Electrode
Collector Insulator
TRACE 1600/1610 Hardware Manual
283