Installing the ECD Exhaust Vent Kit

Installing the ECD Exhaust Vent Kit
452
TRACE 1600/1610 Hardware Manual
The ECD Exhaust Vent Kit (PN 19070024) is used to vent the exhaust gases to a fume hood
(or other exhaust system) when toxic compounds are analyzed.
The kit consists of a fitting M5, an Aluminum washer and a silicone rubber tubing.
To install the ECD exhaust vent kit
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Close the gas supplies.
4. Open the module flap cover.
5. Using a 1/4-in. open-end wrench unscrew and remove the chimney and its washer.
See
Figure
450.
Figure 450. ECD Exhaust Vent Kit Installation (1)
Chimney
Washer