12. Configure and enable the Auxiliary Gas system through the user interface of your GC, or
through the CDS in use. Refer to the TRACE 1600/1610 User Guide.
a. Configuration and enabling through the touch screen.
i.
In the main menu select the Instrument control icon. The Instrument Control
menu appears.
ii. In the Instrument Control menu, select the Auxiliary icon to open the relevant
submenu.
iii. Set the Aux Gas Pressure values as required, then return to main menu.
b. Configuration and enabling through the Chromatography Data System.
i.
In the Configuration window select the Auxiliary tab.
ii. Select the check box Auxiliary Carrier Module 1/2 according to the auxiliary
carrier module installed on your GC.
iii. Select the Auxiliary Pressure check box to enable up to six auxiliary Pressures,
and the adjacent field.
iv. According the inlet ports connected to the Auxiliary Gas Interface, select the
corresponding check box and set the pressure in the adjacent field.
13. Set the normal detector, injector, and GC working conditions.