Adding Modules

Adding an Analog Output Interface

Adding an Analog Output Interface
528
TRACE 1600/1610 Hardware Manual
This section provides instructions for installing the Analog Output Interface (AOI) on your
TRACE 1600/1610 GC.
WARNING This operation must be carried out by authorized and trained Thermo Fisher
Scientific Service Field Engineers.
To install an analog output interface
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
1. Remove the cover of the external modules housing where installing the Analog Output
Interface module. See
Figure 538. Housing Cover Removal
a. Using a T20 Torxhead screwdriver, unscrew and remove the left or right housing
cover screws, then remove the cover from the housing.
2. Install the Analog Output Interface into the housing.
a. Loosen the two hexagonal screws under the module. See
Figure
538.
Left Cover
Screws
Right Cover
Screws
Left/Right Housing Covers
Figure
539.