8
Installing Optional Kits
Installing the Purge & Trap Adapter Kit on the SSL/SSLBKF Injector
354
TRACE 1600/1610 Hardware Manual
b. Place the injector module in its seat. Be sure to insert the 25-pin male connector, on
the bottom of the module, into the 25-pin female connector on the injector seat of
the injector housing.
c. Use a T20 Torxhead screwdriver to tighten the three captive fixing screws without
overtighten.
IMPORTANT To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
10. Remove the top parts of the injector. See
Figure 333. Injector Top Parts Removal
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
a. Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c. Remove the septum holder/liner cap with septum from the injector body head.
11. Replace the liner and the O-ring.
IMPORTANT We suggest replacing the liner currently installed in the injector with the
HS/SPME liner (PN 453A1335). The O-ring currently installed on the liner must be
replaced with the SSL O-ring for the Purge & Trap Adapter (PN MI-290AA1-0001).
a. Use tweezers to remove the current liner with the liner seal (O-ring) from the injector.
CAUTION Be careful not to break the glass liner when removing it.
Glass splinters could fall into the lower part of the vaporization chamber.
Figure
333.
Thermo Scientific