Thermo Scientific TRACE 1600 Hardware Manual page 432

Gas chromatographs
Table of Contents

Advertisement

8
Installing Optional Kits
Installing the NoVent Microfluidics
400
TRACE 1600/1610 Hardware Manual
Figure 373. Location of Detector Modules on your TRACE 1600/1610 GC
Dummy Front Detector Module
CAUTION The gas connection for a dummy module is blocked by a plug.
ATTENTION Le raccordement au gaz d'un module factice est obturé par un bouchon.
To connect the NoVent Microfluidics module to the TRACE 1600/1610 GC
1. Cool the oven, injector or injectors, transfer line, ion source, and any installed GC
detectors to room temperature and shut down the GC.
WARNING - BURN HAZARD: The injectors, detectors, oven, and transfer line may be hot.
Allow them to cool to room temperature before touching them.
2. Push down on the power switch to power off the GC.
3. Shut down the TSQ 8000 Series or ISQ mass spectrometer using the software. See your
mass spectrometer's user guide for more information. The heaters and the turbomolecular
pump power off.
4. If you are using hydrogen as a carrier gas, unscrew the hydrogen safety screw on the front
door of the mass spectrometer.
WARNING - FIRE HAZARD: If you are using hydrogen, do NOT reach over the top of the
instrument to power it off. Instead, reach around the right side or go to the back of the
instrument and flip down the power switch.
Dummy Back Detector Module
Thermo Scientific