Installing Optional Kits
Performing the Dual FPD Detector Configuration
438
TRACE 1600/1610 Hardware Manual
8. Remove the filter a support and the flange assembly from the cell of the secondary FPD
detector.
a. Loosen the three smaller Allen screws that fix the flange to the detector body and
remove it with the spacer that should remain inserted in the flange See
Figure 427. Filter Support and Flange Assembly Removal
b. Now you could access the flame-side heat shield, and the relevant graphite seal.
Loosen the Allen screws that fix the mirror plug and remove it. See
Figure 428. Mirror Plug Removal
c. Insert the handle of a screwdriver or other un-sharpened tool in the combustion
chamber and push the flame-side heat shield and its graphite seal out from its
housing. Act gently to avoid breaking of the heat shield. See
CAUTION While pushing out the heat shield, pay attention not to damage the ignition coil.
ATTENTION Pendant l'extraction de l'écran thermique, veillez à ne pas endommager la
bobine d'allumage.
Fixing Screws
Figure
Figure
429.
Figure
427.
428.