9
Adding Modules
Adding a Generic Detector Interface
524
TRACE 1600/1610 Hardware Manual
Figure 533. Detector Gas Tubing Connection to the GDI Electrical Interface - External View (2)
Note The bending of the gas tubing is indicative.
4. Connect the supply gas to the GDI electrical interface.
Note Use the 1/8-in. Swagelok fittings provided on the gas inlet ports to connect the
gas lines.
a. Connect the gas line to the corresponding inlet port using the appropriate nut and
ferrules. Use a 7/16-in. wrench for tightening the fittings.
b. Repeat
step a
until all the gas lines are connected to the corresponding inlet port on
the GDI electrical interface. See
Figure 534. Gas Lines Connection to the GDI Electrical Interface
Gas 1
IMPORTANT The maximum nominal inlet pressure for all the inputs is 1050 kPa (152 psig),
as indicated on the label under the gas inlets ports on the back of the GC. The working inlet
pressure range is from 400 kPa (58 psig) to 1050 kPa (152 psig).
Figure
534.
Gas 2
Gas 3
Gas 3
Gas 2
Gas 1
Gas 3
Gas 2
Gas 1
Thermo Scientific