Connecting a GSV Backflush System

Connecting a GSV Backflush System
396
TRACE 1600/1610 Hardware Manual
b. Cut 2–3 cm from the column end.
c. Insert the column into the detector base body and slide the ferrule up to the detector
base.
d. Finger-tighten the column retaining nut until it starts to grip the column.
e. Push the column through the detector base according to the detector in use.
f.
Use the 6 mm wrench to tighten the retaining nut. Use enough pressure necessary to
obtain a good seal without overtighten.
To install of a second backflush system
If a second SSL/PTV Backflush system is installed, you can connect both the systems by using
two Tee connectors mounted on the same mounting bracket as schematically shown in
Figure
371.
Figure 371. Schematically Connection of a Second Backflush System
BackFlush Line
Backflush Line
Tip Turn the Tee connector as required to allow the installation of the two precolumns.
This sections provides instructions for connecting your Gas Sampling Valve module with
precolumn and analytical capillary column into the oven of the TRACE 1600/1610 GC.
Refer to the chapters Gas Sampling Valve (GSV) Module in the TRACE 1600/1610 User
Guide.
The Gas Sampling Valve backflush system is provided with the kit PN 19050764.