Instrument Start-up

Connecting TriPlus 500 HS, TriPlus RSH SMART, or TriPlus 100 Autosampler

Thermo Scientific
1. Connect the Mini Delta port on the AI/AS 1610 to the relevant Mini Delta port on the
Trace 1600 Series GC.
2. Turn on the Trace 1600 Series GC and switch on the AI 1610 autosampler.
3. The AI/AS 1610 will automatically run a self-test routine and carry out the following
automatic checks and settings:
• alignment between AI/AS 1610 and the GC injector
• recognize the installed sample tray
• calculation of the syringe zero
Note The self-test routine is carried out every time the safety door on the turret is
closed.
To connect the TriPlus 500 HS to a TRACE 1600/1610
1. Connect the Mini Delta connector to the BACK AS Mini Delta connector on the GC.
2. Connect the 9-pin connector to the GC connector on the TriPlus 500 HS.
Figure 39. TriPlus 500 HS Connection to TRACE 1600/1610
GC back panel
TriPlus 500 HS
back panel
1
Installing the Autosampler
TRACE 1600/1610 Hardware Manual
Installation
35