CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
16. Close the module flap cover.
17. If present, move the autosampler towards the module to restore the original alignment.
18. Open the gas supplies.
19. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
20. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
21. Set the normal oven, injectors, and detectors working conditions.
6
Injectors Advanced Maintenance
Cleaning the PTV Injector Head Assembly
TRACE 1600/1610 Hardware Manual