Page 3
Thermo Fisher Scientific or one of its authorized representatives. TSQ Quantum XLS USA and European Union Regulatory Compliance requirements concerning product safety and electromagnetic compliance of this product are clearly stated in the hard copy of the Declaration of Conformity included with this system.
Page 4
replacement parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific or one of its authorized representatives. FCC Compliance Statement THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL INTErfERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTErfERENCE RECEIVED, INCLUDING INTErfERENCE THAT MAY CAUSE UNDESIRED OPERATION.
Page 5
Request Form Dear User: Read the TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide, and then print and complete the following installation request form. After all items on the form are fulfilled, sign and date the form. Then, mail or fax this form to your local Thermo Fisher Scientific sales/service office. The address and fax number for your local office are located on the following pages.
Page 7
Offices for Thermo Fisher Scientific San Jose Products North America Europe Northeastern Region Austria 265 Davidson Avenue, Suite 101 Wehlistrasse 27b Somerset, NJ 08873 A-1200 Wien Phone......[1] (732) 627-0220 Phone ......[43] (01) 333 50 34-0 Fax ......[1] (732) 627-0260 Fax ......[43] (01) 333 50 34-26 Southern Region Belgium 665 Molly Lane, Suite 140...
Page 8
Offices for Thermo Fisher Scientific San Jose Products - Continued Europe - Continued Australasia and Asia Spain Australia Acer 30-32 P.O. Box 239 Rydalmere Edificio Sertram – Planta 2, Modulo 3 Unit 14, 38 - 46 South Street ES-08038 Barcelona Rydalmere, N.S.W.
Page 9
WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union (EU) Member State, and these companies should dispose of or recycle this product.
Page 10
Conformité DEEE Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant: Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devrait être collecté...
Preface The Thermo Scientific TSQ Quantum XLS™ and TSQ Quantum GC™ systems are member of the TSQ™ Quantum family of mass spectrometers. This Preinstallation Requirements Guide provides information to assist in planning and preparing your lab site for the system prior to delivery and installation. Read each section carefully to be sure that your laboratory is ready for the system installation.
Knowledge base www.thermokb.com Find software updates and utilities to download at mssupport.thermo.com. To contact Customer Service for ordering information Phone 800-532-4752 561-688-8731 E-mail us.customer-support.analyze@thermofisher.com Web site www.thermo.com/ms TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide Thermo Scientific...
Page 15
To suggest changes to documentation or to Help • Fill out a reader survey online at http://www.surveymonkey.com/s.aspx?sm=R7gKOvhLXn3NTkpK2BefHQ_3d_3d. • Send an e-mail message to the Technical Publications Editor at techpubs-lcms@thermofisher.com. Thermo Scientific TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide...
For additional information, request specific preinstallation support directly through your local Thermo Fisher Scientific office. Thermo Scientific TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide...
1 × 1.22 m (3 × 4 ft) and be capable of supporting the weight of the data system computer, monitor, and optional printer [26 kg (57 lbs)]. Telephone Install a telephone line near the workbench. Thermo Scientific TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide...
102 cm [40 in.], w 81 cm [32 in.], h 43 cm [17 in.], weight: 46 kg [100 lb]), computer, forepump, monitor, and other options—are shipped in their own containers. Their dimensions and weights are less than that of the container for the TSQ Quantum XLS mass spectrometer.
Page 21
Table 2 lists the space requirements and weights of the typical TSQ Quantum XLS or TSQ Quantum GC system components. Table 2. TSQ Quantum XLS and TSQ Quantum GC system dimensions and weights...
Page 22
Figure 1. TSQ Quantum XLS and TSQ Quantum GC systems space requirements Table Top Table Top > 173 cm 122 cm (68 in.) (48 in.) 137 cm (54 in.) 48 cm 41 cm 19 cm 56 cm (19 in) (16 in.) (7 in.)
Page 23
36 cm (14 in.) during the installation or removal of the transfer line. See Figure Figure 2. Top view and dimensions of the TRACE GC Ultra gas chromatograph (left) and TSQ Quantum XLS or TSQ Quantum GC mass spectrometer (right) 173 cm (68 in.)
CAUTION Your instrument is designed to work in a controlled electromagnetic environment. Do not use radio frequency transmitters, such as mobile phones, in close proximity to the instrument. TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide Thermo Scientific...
Operating Environment Attention to the operating environment will ensure continued high performance of your TSQ Quantum XLS or TSQ Quantum GC system. Any expenditures for air conditioning are more than offset by good sample throughput and reduced repair costs. Table 3 lists the operating environment requirements and indicates where you will find additional details.
5 °C or more over a five-minute period of time can affect performance. The air conditioning load for a basic TSQ Quantum XLS or TSQ Quantum GC system (with a Trace GC Ultra gas chromatograph) is approximately 5470 W (19 000 Btu/h).
Maintain the relative humidity of the operating environment between 40 and 80 percent, with no condensation. Operating a TSQ Quantum XLS or TSQ Quantum GC system in an environment with very low humidity can cause the accumulation and discharge of static electricity that can shorten the life of the electronic components.
Electrostatic discharge (ESD) can damage the electronic components of your TSQ Quantum XLS or TSQ Quantum GC system. TSQ Quantum XLS and TSQ Quantum GC instruments are designed to withstand ESD up to 15 kV (air discharge) and 8 kV (contact discharge) with all panels in place. However, removing the panels and handling the PCBs without proper precautions might damage the electrical components or cause them to fail prematurely.
50 μs. Power Monitoring Devices Before connecting your TSQ Quantum XLS or TSQ Quantum GC mass spectrometer to its line power, Thermo Fisher Scientific recommends that you monitor the line power 24 hours a day for seven consecutive days.
Page 30
30 A (230 V ac). • Earth ground hardwired to the main panel. Connecting the TSQ Quantum XLS or TSQ Quantum GC Mass Spectrometer, GC, and Other Modules to Wall Outlets Balance the current load on the circuits that your system is connected to.
Thermo Fisher Scientific. Quality of Power The quality of power supplied to your TSQ Quantum XLS or TSQ Quantum GC system is very important. The line voltage must be stable and within the specifications listed in this guide. The line voltage must be free of fluctuations due to slow changes in the average voltage, surges, sags, or transients.
Thermo Fisher Scientific does not endorse any power monitoring company, nor does it endorse products other than its own. Companies and products listed in this guide are given as examples only. TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide Thermo Scientific...
Available Outlets Available Outlets The TSQ Quantum XLS and TSQ Quantum GC mass spectrometers are designed to operate at a nominal voltage of 230 V ac, 50/60 Hz. Line voltages can vary between a minimum of 207 V ac and a maximum of 253 V ac.
Page 34
6 shows the optimum location of the power outlets. The power cable from the TSQ Quantum XLS system is 3 m (9 ft) long. The cables from the personal computer, monitor, and printer are approximately 2 m (6 ft) long.
Line Power Connecting the TSQ Quantum XLS or TSQ Quantum GC Mass Spectrometer, GC, and Other Modules to Wall Outlets Installing a complete GC/MS system can require extensive electrical resources. Plan the power system properly, with numerous outlets, to ensure that you can connect and power all of your equipment.
Technical Assistance Occasionally, you might encounter line power sources of unacceptable quality that adversely affect the operation of TSQ Quantum XLS and TSQ Quantum GC systems. You are responsible for correcting line power problems. Contact your Thermo Fisher Scientific office for assistance in monitoring the line voltage in your laboratory and in selecting a line conditioner.
Waste and Exhaust The arrangement of the exhaust system for the TSQ Quantum XLS or TSQ Quantum GC instrument can affect instrument performance. Also, take care to collect and dispose of solvent waste properly. You are responsible for making the proper exhaust arrangement that is required to operate your system.
3 L/min and a continuous flow rate of 1 L/min. Solvent Waste Consult local regulations for the proper method of disposing solvent waste. TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide Thermo Scientific...
The regulator output pressures must be consistent with the pressures listed in this table. Equip each regulator with a 1/8 in. Swagelok™ compression fitting or equivalent to connect a 1/8 in. metal tube. Thermo Scientific TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide...
The regulator output pressures must be consistent with the pressures listed in Table 10. Each regulator must be equipped with a 1/8 in. Swagelok compression fitting (or equivalent) to connect a 1/8 in. metal tube. TSQ Quantum XLS and TSQ Quantum GC Preinstallation Requirements Guide Thermo Scientific...
Installation Prior to having your TSQ Quantum XLS or TSQ Quantum GC system installed, make sure to complete all preparations described in the previous chapters. After you have completed your laboratory site preparation, mailed or faxed the TSQ Quantum XLS or TSQ Quantum GC Installation Request Form, and received the system, call your Thermo Fisher Scientific office to set up an installation date.
Verify that your lab meets the following preinstallation requirements before the Thermo Fisher Scientific service engineer installs the TSQ Quantum XLS or TSQ Quantum GC system. To check off each requirement, use the TSQ Quantum XLS and TSQ Quantum GC Installation Request Form at the front of this guide.
20. The principal operator is available during the installation and certification period. Installation Kits The following kits are shipped with the TSQ Quantum XLS and TSQ Quantum GC systems: • MS Ship Kit (P/N 70111-62033): Contains installation components, such as the exhaust tubing and power cords.
Installation Installation Installation Once your new TSQ Quantum XLS or TSQ Quantum GC system is on site and ready for installation, a Thermo Fisher Scientific field service engineer installs it. During the installation, the field service engineer demonstrates the following: •...
Instrument Arrival Electronic equipment carriers that specialize in the handling of delicate machinery ship the TSQ Quantum XLS or TSQ Quantum GC system to your site. Occasionally, however, equipment is damaged in transit. Take the following precautions when receiving material: •...