Installing the Purge & Trap Adapter Kit on the SSL/SSLBKF Injector
Installing the Purge & Trap Adapter Kit on the SSL/SSLBKF Injector
352
TRACE 1600/1610 Hardware Manual
This section provides the instruction for installing the Purge & Trap Adapter kit
(PN 19050730) on the SSL/SSLBKF injector for the use with a Purge & Trap system.
See
Figure
331.
Figure 331. Purge & Trap Adapter
Purge and Trap Adapter
To install the Purge & Trap Adapter
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Close the gas supplies.
4. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. Remove the column end from the injector.
a. Open the front door of the GC.