Installing Optional Kits
Installing the Packed Column Adapters
360
TRACE 1600/1610 Hardware Manual
5. Remove the bottom parts of the SSL injector. See
Figure 340. SSL Injector Bottom Parts Removal
a. Using the 1/2 wrench provided, unscrew the retaining nut with the base seal and the
washer from the bottom of the injector.
CAUTION Make sure that the liner does not come out from the bottom of the
injector. Save the bottom parts of the SSL injector in a safe place because they will
be reused when you restore the original configuration for capillary columns.
ATTENTION Assurez-vous que le liner ne sorte pas par le bas de l'injecteur.
Conservez les pièces inférieures de l'injecteur SSL en lieu sûr, car elles seront
réutilisées pendant la restauration de la configuration d'origine pour les colonnes
capillaires.
6. Install the adapter on the bottom the injector interposing the silver seal. See
Figure
340.
SSL Injector Bottom
Base Seal
Washer
Retaining Nut
Figure
341.
Thermo Scientific