Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector with Backflush (SSLBKF)
98
TRACE 1600/1610 Hardware Manual
Figure 84. SSLBKF Injector Components
Septum Cap
Septum
Ring Nut
Septum Holder/Liner Cap
Liner Seal
Liner
Body Head External O-ring
Body Head Internal O-ring
Base Seal
Washer
Retaining Nut
(Terminal Fitting for Capillary Column)
Backflush Line Fitting
Ferrule
Fitting Cap, Backflush Line
The SSLBKF injector periodic maintenance includes:
• Replacing the SSLBKF septum
Replace the septum at least after every 200 injections, or every time a problem related to
septum damage, or wear occurs.
See
"Replacing the SSLBKF Septum"
• Cleaning or replacing the SSLBKF liner
Check the liner for contaminants, debris, breakage, and proper installation. The liner
must be replaced depending on the number of injections performed, and the
characteristics of the samples injected. Typical symptoms will indicate that the liner must
be replaced. The most common is the appearance of tailing peaks in the chromatogram,
particularly for polar compounds.
See
"Cleaning or Replacing the SSLBKF Glass Liner"
on
page
100.