c. If your GC is equipped with the touch screen as user interface, select Back or Front
Column icon in the Configuration menu, otherwise perform the Column
Evaluation through the Chromatography Data System by selecting the proper
function.
d. Select the column and input the physical characteristics of the column.
e. If a pre-/post-column is present, set the length and nominal internal diameter of the
pre-/post-column in the same valid ranges for the column. The following two lines
are added to the menu.
f.
According to the physical characteristics of the column, the system calculates and
displays the relevant Column K-factor.
Note For the most reproducible results, you should conduct a more detailed
column evaluation. However, the following steps, while recommended, are not
required.
g. Start column evaluation. At the end of the routine, a message will indicate that
evaluation was successful.
h. Expect a K-factor of approximately 0.7 – 0.9 for a 15 m, 0.25 mm ID column
(1.3 – 2.0 for a 30 m, 0.25 mm ID column). If the column does not report a K-factor
within this range or within 0.1 units of the previous stored value, check for a leak or
broken column using the leak detector. The K-factor is a measured resistance for the
column. A K-factor that is too low may indicate a leak in the system, while a K-factor
that is too high may indicate a blockage.
23. Disconnect the column-flowmeter.
a. Disconnect the column from the column-flowmeter connector.
b. Remove the clear plastic component, including its fittings, from the oven and set it
aside.
c. Close the GC door.
24. Condition the column.
The column must be conditioned before inserting it into the detector.
Column conditioning consists of passing a carrier gas flow through the column heated at
a programmed temperatures as described in the column manufacturer's instructions.
In case the column does not have any column conditioning instructions, perform the
column conditioning by setting a final temperature up to 10 °C - 20 °C below its
recommended maximum temperature.
8
Installing Optional Kits
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
TRACE 1600/1610 Hardware Manual
373