Performing Injectors Routine Maintenance
Maintaining a Programmable Temperature Vaporizing Injector with Backflush (PTVBKF)
138
TRACE 1600/1610 Hardware Manual
Figure 110. PTVBKF Injector Components
Septum Cap
Septum
Liner Cap
Liner Seal
Liner
Silver Seal
Terminal Fitting for Capillary Column
Ferrule
Split Nut
Backflush Line
Backflush Line Fitting
Ferrule
Fitting Cap, Backflush Line
The PTVBKF injector periodic maintenance includes:
• Replacing the PTVBKF septum
Replace the septum at least after every 200 injections, or every time a problem related to
septum damage, or wear occurs.
See
"Replacing the PTVBKF Septum"
on
page
140.