b. Place the injector module in its seat. Be sure to insert the 25-pin male connector, on
the bottom of the module, into the 25-pin female connector on the injector seat of
the injector housing.
c. Use a T20 Torxhead screwdriver to tighten the three captive fixing screws without
overtighten.
IMPORTANT To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
11. Remove the top parts of the injector. See
Figure 413. Injector Top Parts Removal
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal (O-ring)
Liner
a. Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c. Remove the septum holder/liner cap with septum from the injector body head.
12. Remove the liner and its seal.
a. Use tweezer to remove the current liner with the liner seal (O-ring) from the injector.
13. Remove the injector body. See
Installing the Hot Injection Adapter Kit on the SSL/SSLBKF Injector
Figure
333.
Figure
414.
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits