Cleaning Or Replacing The Glass Liner - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

3
Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)

Cleaning or Replacing the Glass Liner

120
TRACE 1600/1610 Hardware Manual
c. Avoid touching the septum with your fingers. Insert a new septum into the septum
holder using tweezers.
d. Screw and tighten the septum cap to finger-tight.
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
8. Close the module flap cover.
9. If present, move the autosampler towards the module to restore the original alignment.
10. Wait two or three minutes for helium to purge the inlet.
11. Turn the inlet flow back on and enable the inlet heater.
12. Set the normal injector, detector, and GC working conditions.
To clean or replace the glass liner
Figure 100. HeS-S/SL Injector: Glass Liner Replacement
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal (O-ring)
Liner
Note Care should be taken when performing the procedure in order to keep from
damaging analytical columns.