Figure 419. Hot Injection Adapter (3)
c. Screw and tighten the septum cap to finger-tight. See
Figure 420. Hot Injection Adapter (4)
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
17. Close the module flap cover.
18. Connect the hot injection adapter to the device for the sampling of gas. See the example
in
Figure
421.
Installing the Hot Injection Adapter Kit on the SSL/SSLBKF Injector
Figure
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits