4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
196
TRACE 1600/1610 Hardware Manual
d. Place the jet on a paper towel and let it air dry. When the jet is dry, insert the jet into
the detector base body and tighten it. See
Figure 166. FPD Jet Reinstallation
Note If after cleaning the physical condition of the jet does not permit its reuse,
replace the jet with a new one.
8. Reinstall the FPD detector on the base.
a. Place the detector on its base body, ensuring that the Aluminum ring has been
inserted in the correct position, then tighten the fixing nut. See
Figure 167. FPD Detector Reinstallation
b. Reconnect the signal, excitation voltage and ignition/heating cables to the detector.
See
Figure
168.
Figure
166.
Figure
167.
FPD Fixing Nut
Aluminum Ring
Thermo Scientific