Installing And Connecting The Auxiliary Gas Module - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

10
Adding Systems
Adding an Auxiliary Gas System

Installing and Connecting the Auxiliary Gas Module

568
TRACE 1600/1610 Hardware Manual
To install and connect the Auxiliary Gas Module
1. Remove the cover of the external modules housing where installing the module. See
Figure
589.
Figure 589. Housing Cover Removal
a. Using a T20 Torxhead screwdriver, unscrew and remove the left or right housing
cover screws.
b. Remove the cover from the housing.
2. Remove the manifold protecting plate
a. Using a T20 Torxhead screwdriver, unscrew the two fixing screws, and remove the
protecting plate from the manifold. Save the protecting plate and the fixing screws.
Figure 590. Manifold Protecting Plate
3. Install the module into the housing
Left Cover
Screws
Right Cover
Screws
Left/Right Housing Covers
Protecting Plate
Fixing Screws
Thermo Scientific