Figure 396. Inserting the Column into the Transfer Line
Ferrule
Transfer Line
18. Tighten the transfer line nut until it is just secure enough so that you cannot move it.
19. Loosen the nut by turning it exactly 1 flat backward.
20. Position the column in the transfer line. Use the septum as a guide to measure the correct
length you should insert the column. Be careful not to change the location of the septum
on the column.
Figure 397. Positioning the Column in the Transfer Line
21. Tighten the nut 1 flat forward—back to where it is secure enough in the transfer line that
you cannot move it.
22. Tighten the nut 1 additional quarter turn.
23. Remove the cut septum.
Transfer Line
Nut
Transfer Line
Septum
8
Installing Optional Kits
Installing the NoVent Microfluidics
Column
Septum
Column
TRACE 1600/1610 Hardware Manual
413