Maintenance Button

Powering On the TRACE 1300/TRACE 1310

Thermo Scientific
When pressed, the Maintenance button allows cooling at 60 °C of the selected heated zones,
which allows the required maintenance. Carrier and detector gases must be closed manually.
This function also performs the following actions automatically:
• turns off the flame and the fuel gases of the FID and FPD detectors.
• turns off the thermionic source, the hydrogen, and the air flows of the NPD detector.
• turns off the filaments of the TCD detector.
To power on the GC
1. Install the GC column (see
2. Open the supply gases.
3. Plug the power cable to the AC Input connector (Main socket) on the GC, and to the
wall outlet. See
Figure
4. Be sure the carrier gas flowing through the column and the detector gases flowing
through the detector.
Figure 70. GC Power On
120 Vac or 230 Vac
Power Supply Label
Power Switch ON Position
AC Input Connector
"Replacing a Column"
70.
TRACE 1300 and TRACE 1310 Hardware Manual
2
Performing Routine Maintenance
Maintenance Button
on
page
73).
69