Thermo Scientific TRACE 1600 Hardware Manual page 153

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Ensure the Enable conservation field in the configuration page of the inlet reads Yes.
4. Check that there is at least 20 mL/min split flow exiting the split line, then turn the
column flow to Off in the GC user interface and allow the inlet to depressurize.
5. Put the autosampler away if present.
6. Open the module flap cover.
7. Remove the top parts of the injector.
a. Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c. Remove the septum holder/liner cap with septum from the injector body head.
8. Remove the liner.
a. Using tweezers, remove the liner with the liner seal (O-ring) from the injector.
9. Replace or clean the liner.
• If you are going to clean the dirty liner, go to
• If you are going to use a new liner, go directly to
10. Clean the liner.
a. Put the liner into an ultrasonic cleaner filled with a methanol/acetone mixture (1:1).
b. Sonicate the liner for about half an hour.
c. Using tweezers, remove the liner from the bath and dry it with compressed clean air.
11. Install the liner.
a. Hold the new (or cleaned) liner with tweezers place a new liner seal over the liner.
b. Insert the liner into the injector, and push it gently towards the bottom of the
injector.
12. Reinstall the top parts of the injector.
a. Place the septum holder/liner cap with the septum on the body head of the injector,
and fix it by screwing the ring nut.
b. Screw and tighten the septum cap to finger-tight.
3
Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
step
10.
step
11.
TRACE 1600/1610 Hardware Manual
121