Figure 104. PTV Injector Components
Septum Cap
Septum
Liner Cap
Liner Seal
Liner
Silver Seal
Terminal Fitting for Capillary Column
Ferrule
Split Nut
The PTV injector periodic maintenance includes:
• Replacing the PTV septum
Replace the septum at least after every 200 injections, or every time a problem related to
septum damage, or wear occurs.
See
"Replacing the PTV Septum"
• Cleaning or replacing the PTV liner
Check the liner for contaminants, debris, breakage and proper installation.
3
Performing Injectors Routine Maintenance
Maintaining a Programmable Temperature Vaporizing Injector (PTV)
on
page
129.
TRACE 1600/1610 Hardware Manual
127