b. Reinstall the analytical column.
14. Reinstall the liner
a. Place the liner seal over the liner, then, Using tweezers, insert the liner into the
injector and push it gently towards the bottom of the injector.
15. Reinstall the top parts of the injector.
a. Place the septum holder/liner cap with the septum on the body head of the injector.
Fix them by screwing the ring nut.
b. Screw and tighten the injector cap.
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
16. Close the module flap cover.
17. If present, move the autosampler towards the module to restore the original alignment.
18. Turn the carrier gas on.
19. Set the normal oven and injector working conditions.
6
Injectors Advanced Maintenance
Cleaning the SSLBKF Injector Body
TRACE 1600/1610 Hardware Manual
255