Thermo Scientific TRACE 1600 Hardware Manual page 374

Gas chromatographs
Table of Contents

Advertisement

7
Detectors Advanced Maintenance
Replacing the FPD Photomultiplier Tube
342
TRACE 1600/1610 Hardware Manual
a. Place the detector on its base body, ensuring that the Aluminum ring has been
inserted in the correct position, then tighten the fixing nut. See
Figure 324. FPD Detector Reinstallation
b. Reconnect the signal, excitation voltage and ignition/heating cables to the detector.
See
Figure
325.
Figure 325. FPD Cables Connection
Figure
324.
Thermo Scientific