Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
204
TRACE 1600/1610 Hardware Manual
CAUTION When removing the covers, pay attention to the heater element and the
temperature probe.
ATTENTION Pendant la dépose / réinstallation des couvercles, faites attention à l'élément
chauffant et à la sonde de température.
a. Using a 5/16-in tube wrench, remove the two screws on the top of the detector body,
and the front and back screws on the lower part of the detector body. See
Figure 177. Detector Body Dismounting (1)
b. Remove the left and right covers of the FPD detector body and the insulating
material. The glow-plug is now accessible. See
Figure 178. Detector Body Dismounting (2)
8. Replace the defective glow-plug. See
Top Screws
Right Cover
Front Screw
Back Screw
Figure
178.
Figure
179.
Figure
177.
Right Cover
Insulating
Material
Insulating
Material
Left Cover
Glow-plug Cable
Terminal body