Detectors Advanced Maintenance
Cleaning or Replacing the NPD Jet
308
TRACE 1600/1610 Hardware Manual
Figure 278. Reinstall Thermionic Source (3)
20. Close the module flap cover.
21. Plug in the signal cable into its contact on the detector module. See
Figure 279. Plug NPD Signal Cable
22. If present, move the autosampler towards the module to restore the original alignment.
23. Plug the power cable to the AC input connector of the NPD Thermionic Source Power
Module and to the wall outlet. The LED marked On blinks green.
If other external modules are present, plug the power cable to the AC Input connector of
each module, and to the wall outlet.
24. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
25. Turn the make-up gas on.
26. Heat the detector the working temperature.
Figure
279.