Thermo Scientific TRACE 1600 Hardware Manual page 227

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 164. FPD Detector Removal
Note Do not lose the Aluminum ring inserted between the detector head and the
base body.
7. Clean the FPD Jet.
a. Using the 5 mm wrench provided with the GC, loosen the jet and remove it from the
detector base body. See
Figure 165. FPD Jet Removal
b. Place the jet in the ultrasonic cleaner filled with liquid detergent and clean it for
about five minutes.
c. Handling the jet with forceps or tweezers, rinse the jet with distilled water, then with
methanol.
4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
Figure
165.
FPD Fixing Nut
Aluminum Ring
TRACE 1600/1610 Hardware Manual
195