Note If you already have a mounting bracket installed in your GC with either the SSL
backflush or PTV backflush system, skip this step.
To install the bracket in the GC oven
1. Attach the bracket to the oven in one of the following areas. See the following figure for
the mounting bracket locations.
Figure 376. Brackets Attached to the Pre-Drilled Holes in the GC Oven
5-port
connection
3-port
connection
2. Using a T20 Torxhead screwdriver, insert both of the fixing screws halfway into the
pre-drilled holes.
Figure 377. Attaching Fixing Screws to the Oven Wall and Floor
3. Slide the bracket until the screws snap into place.
8
Installing Optional Kits
Installing the NoVent Microfluidics
GC Oven
TRACE 1600/1610 Hardware Manual
403