Installing the Column the First Time

Thermo Scientific
Figure 56. Column Rack
Button-holes
Oven Heater
Baffle
To install a column rack
1. Open the front door of the GC.
2. Slightly press the sides of the column rack and insert the two hooks into the
corresponding button-holes on the oven heater baffle.
This section contains instruction for installing the capillary column into the GC oven, for
connecting the column ends to the injector and the detector and for carrying out the column
conditioning, the leak test and the column evaluation.
CAUTION The injector and detector modules are shipped with a plug. Before installing
the column, remove the plug from the injector and detector modules. Unscrew the
column retaining nut from the bottom of the injector and the detector modules, then
remove the blind ferrule.
ATTENTION Les modules d'injecteur et de détecteur sont fournis avec un bouchon. Avant
d'installer la colonne, retirez le bouchon des modules d'injecteur et de détecteur. Dévissez
l'écrou de fixation de la colonne du bas des modules d'injecteur et de détecteur, puis
retirez la ferrule aveugle.
1