To install the hot injection adapter on the injector
1. Assemble the hot injection adapter if not already done. See
Figure 411. Adapter Assembly
Liner Seal
Liner
Stem
a. Place the liner seal on the liner.
b. Insert the liner into the stem.
c. Couple the head and the stem paying attention to properly align the corresponding
holes doing match the polarizing guides on the head and on the flange of the stem.
d. Fix the stem to the head screwing the three fixing screws using a T8 Torxhead
screwdriver.
Note The stem conducts heat from the body of the SSL/SSLBKF injector to the
head of the adapter allowing a hot injection.
2. Put the GC in standby condition.
3. Cool the oven, injector, and detector to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
4. Close the gas supplies.
5. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
6. Remove the column end from the injector.
Installing the Hot Injection Adapter Kit on the SSL/SSLBKF Injector
Figure
Guides
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits
411.
Fixing
Screws
425