Figure 77. SSL Injector: Components
Septum Cap
Septum
Ring Nut
Septum Holder/Liner Cap
Liner Seal
Liner
Body Head External O-ring
Body Head Internal O-ring
Base Seal
Washer
Retaining Nut
Terminal Fitting for Capillary Column
The SSL injector periodic maintenance includes:
• Replacing the SSL septum
Replace the septum at least after every 200 injections, or every time a problem related to
septum damage, or wear occurs.
See
"Replacing the SSL Septum"
• Cleaning or replacing the SSL liner
Check the liner for contaminants, debris, breakage, and proper installation.
The liner must be replaced depending on the number of injections performed, and the
characteristics of the samples injected. Typical symptoms will indicate that the liner must
be replaced. The most common is the appearance of tailing peaks in the chromatogram,
particularly for polar compounds.
See
"Cleaning or Replacing the SSL Glass Liner"
Tip It is good practice to replace the septum every time you replace the glass liner.
3
Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector (SSL)
Bottom Fittings
on
page
87.
on
page
89.