Thermo Scientific TRACE 1600 Hardware Manual page 57

Gas chromatographs
Table of Contents

Advertisement

Thermo Scientific
Figure 27. Left Side Panel Fixing Screw
c. Slide the panel towards the back of the instrument up to the stop.
d. Remove the panel by pulling it outward. Be aware that the ground wire is attached to
the panel.
2. Prepare the duct for the ISQ Series or TSQ Series transfer line inner tube.
a. Remove the partial cut plate on the left panel for that corresponds to your MS.
See
Figure
28.
Figure 28. Left Panel for MS Partial Cut Plate
b. On the exterior wall of the oven box, remove the pre-shaped plug of insulating
material from the duct provided. See
Coupling to a Mass Spectrometer
Duct for ISQ Series or
TSQ Series Partial Cut Plate
Figure
29.
TRACE 1600/1610 Hardware Manual
1
Installation
25