Configuring The Data System; Column Installation Requirements; Using The Correct Fittings - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

Configuring the Data System

Column Installation Requirements

Using the Correct Fittings

Thermo Scientific
You can use the self-adhesive labels found in the standard outfit of the Data System to note
the IP address of the instrument. Stick the label where it can be easily found when configuring
the data system.
1. Launch the Data System.
2. Enter the IP address of the GC.
IMPORTANT If you connect through hubs over a 10 Mbit/s network, it is recommended
that no more than five GC units be connected on the same network trunk. If you have a
switched network, this does not apply.
The column is where the separation takes place. It should be positioned inside the oven on the
column rack. The column ends should align correctly with the injector and the detector
bottom fittings. The GC oven controller accurately controls the column temperature.
Each column has a maximum recommended operating temperature. To protect the column
from excessively high temperatures, remember to set the Maximum temperature parameter
for the oven.
Note For detailed information on maximum operating temperature, refer to the column
manufacturer's instructions provided with the column.
Related Topics
"Using the Correct Fittings"
"Installing the Adapters for Encapsulated Graphite Ferrules"
"Installing the Column Rack"
To connect a capillary column to the injector and detector base body, you must use the proper
column ferrules and retaining nuts.
• Retaining Nut — Dedicated retaining nuts are required to connect capillary columns to
injector and detector base bodies, and the Tee connector. See
on
page 45
on
page 52
1
Column Installation Requirements
on
page 48
Table
3.
TRACE 1600/1610 Hardware Manual
Installation
45

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trace 1610

Table of Contents