Installing Optional Kits
Installing the NoVent Microfluidics
Before replacing the column or performing maintenance on an injector module, flip the
switch to the On position. Ensure the mass spectrometer is in standby mode and the filament
is off. The status LED lights up as solid green indicating that the solenoid valve is activated.
The carrier gas flows into the mass spectrometer to avoid breaking vacuum. See
Figure
405.
Figure 405. Module in On Condition
Status LED On
Switch is in the On Position
When you have completed your tasks, flip the switch to the Off position. The status LED
turns Off indicating the solenoid valve is deactivated.
Thermo Scientific
TRACE 1600/1610 Hardware Manual
419