Thermo Scientific TRACE 1600 Hardware Manual page 274

Gas chromatographs
Table of Contents

Advertisement

5
GC Main Frame Advanced Maintenance
Replacing the Flap Motor
242
TRACE 1600/1610 Hardware Manual
13. Reconnect the power plug to connector J10 on the backplane board.
14. Reinstall the back cover.
a. Reconnect the ground wire to the back cover terminal.
b. Replace the cover proceeding in the reverse order in which it was removed.
15. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
16. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
17. Set the normal injector, detector and GC working conditions.