Installing The Adapters For Encapsulated Graphite Ferrules - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

1
Installation
Column Installation Requirements

Installing the Adapters for Encapsulated Graphite Ferrules

48
TRACE 1600/1610 Hardware Manual
CAUTION Using the wrong size ferrules causes leaks and contamination.
ATTENTION L'utilisation de ferrules de taille appropriée entraînera des problèmes de fuite
et de contamination.
When connecting columns, consider that overtightening compression ferrules does not
necessarily produce a stronger, leak-free joint. In fact, very often the reverse is true. Too
much pressure can cause a leak in the joint and make it very difficult to reseal the joint
when changing columns.
This section provides instructions for installing the optional adapters for connecting the
capillary column to the SSL injector and FID, NPD, TCD, ECD, and FPD detector by using
the encapsulated graphite ferrules. The adapters are provided with the kit PN 19050759.
The kit includes the parts listed in
Table 8. Adapters and 1/2-in. Wrench
Part
To install the adapters for encapsulated graphite ferrules
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
Table
8.
Description
Adapter for connection to the SSL injector
Adapter for connection to the FID, NPD, ECD, TCD, or
FPD detector
1/2-in. wrench