b. Connect the touch screen and ground cables to the proper connection points located
on the front of the door. See
Figure 614. Cables Connections
8. Mount the TRACE 1610 cover door proceeding in the reverse order in which the
TRACE 1600 cover door was removed.
9. Remount the door handle.
10. Open the gas supplies.
11. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
12. Power on the GC.
a. Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
13. Configure the system.
14. Set the normal detector, injector, and GC working conditions.
Upgrading a TRACE 1600 to a TRACE 1610
Figure
614.
TRACE 1600/1610 Hardware Manual
11
Upgrading Equipment
589