6. Replace the chimney and its washer with the washer and the fitting M5 provided in the
kit. Then screw the fitting M5 using the 1/4-in open-end wrench. See
Figure 451. ECD Exhaust Vent Kit Installation (2)
Fitting M5
Washer
7. Close the module flap cover.
8. Connect an end of the silicone rubber tubing provided on the fitting M5, then guide the
other end to the fume hood. See
Figure 452. ECD Exhaust Vent Kit Installation (3)
To Fume Hood
(or other exhaust
system)
9. Open the gas supplies.
10. Set the normal detector working conditions.
Figure
452.
TRACE 1600/1610 Hardware Manual
8
Installing Optional Kits
Installing the ECD Exhaust Vent Kit