Installation
Connecting the PTV/PTVBKF Cryogenic System
20
TRACE 1600/1610 Hardware Manual
Figure 19. PTV Cryo System with CO
Carbon dioxide must be supplied by a high-pressure cylinder with a dip tube. It is your
responsibility to ensure the delivery connection from the carbon dioxide cryogenic supply is
adaptable to 1/8-in. tubing.
Figure 20. Carbon Dioxide Container
CORRECT
Dip tube
CAUTION High pressures and extremely low temperatures make pressurized CO
hazardous material. High concentrations of CO
follow the safety precautions and delivery system design recommended by your gas
supplier.
ATTENTION Le CO
sous pression est dangereux en cas de pression élevée ou de
2
température extrêmement basse. Le CO
d'éviter toute blessure, respectez les consignes de sécurité et la conception de système
d'alimentation de votre fournisseur de gaz.
2
Figure 20
shows the proper carbon dioxide container configuration.
WRONG
are dangerous. To avoid injury, always
2
est dangereux à haute concentration. Afin
2
a
2