Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)

Replacing the Carrier and Split Lines Filters

122
TRACE 1600/1610 Hardware Manual
CAUTION Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
ATTENTION Ne serrez pas excessivement le capuchon du septum, car cela pourrait
endommager le septum et affecter les performances de l'instrument.
13. Close the module flap cover.
14. If present, move the autosampler towards the module to restore the original alignment.
15. Wait two or three minutes for helium to purge the inlet.
16. Turn the inlet flow back on and enable the inlet heater.
17. Set the normal injector, detector, and GC working conditions.
CAUTION The dimensions of the filters are different. The filter on the split gas line is
bigger than the filter on the carrier gas line. Do not invert their position when you replace
them. It is not necessary to replace the filters together.
ATTENTION Les dimensions des filtres sont différentes. Le filtre sur la conduite de gaz à
débit divisé est plus grand que le filtre sur la conduite de gaz vecteur. N'inversez pas leur
position pendant leur remplacement. Il n'est pas nécessaire de remplacer les filtres en
même temps.
To replace the active carbon filters on carrier gas line and split line
Thermo Scientific